Smoke Density Chamber

Short Description:

This instrument has been established for many years and is used widely in all industrial sectors for the determination of smoke generated by solid materials and assemblies mounted in the vertical position within a closed chamber.​


  • Min.Order Quantity: 1 unit
  • Price: Negotiable
  • Package : Plywood Case
  • Model: SL838
  • Standards: AITM 2.0007A,AITM 2.0007B,AITM 3.0005、BSS 7238,BSS 7239、FAR PART 25 APPENDIX F Part V
  • Product Detail

    Product Tags

    Smoke Density Chamber Electronic ignition system with PID temperature controller Analysis of various gases

    Product Profile of Smoke Density Chamber

    This instrument has been established for many years and is used widely in all industrial sectors for the determination of smoke generated by solid materials and assemblies mounted in the vertical position within a closed chamber.​

    Product parameters of Smoke Density Chamber

     Size  1560 mm (W) x 2220 mm x (H) x 1060 mm
     Weight  190KG
     Standards  AITM 2.0007A,AITM 2.0007B,AITM 3.0005、BSS 7238,BSS 7239、FAR PART 25 APPENDIX F Part V
     Application  Aviation 
     Electrical  110V AC 60Hz / 230V AC 50Hz, 15A
     Ambient Temperature  Operating 10°C to 35°C

     

    Product Feature of Smoke Density Chamber

    Powder coated liner removable for repair/replacement. Powder coated exterior.
    Large tempered double glazed door for easy viewing and cleaning.
    PID temperature controller for constant furnace setting.
    Coil type tubular radiant heater with integrated thermocouple. Master electrical shut-off switch with maintenance lockout. Pneumatic actuated test sample transport system for repeatability of tests.
    Electronic ignition system.
    Calibrated flowmeters Propane and air flow measurement. Master gas on/off control with safety latching circuit to prevent gas flow following power interruption.
    2 removable 6 flame pilot burners for normal test and wire test. 4 ports on roof of chamber for toxicity analysis.
    3 stainless steel sample holders complete with backer board, spring and spring pin.
    1 blank sample for wall protection. Wire sample holder
    Calibration system including 1 water cooled heat flux gauge (0-5 W/cm2), holder and integral cooling system.
    3 optical filters for linearity calibration.
    Furnace location and pilot burner location tools. Pilot burner cleaning tool.
    Customized data acquisition software and hardware for calibration and testing with USB interface for connection to external computer (administration rights required for computer)
    Data stored in csv format for easy exporting to spreadsheet programs.

    Spare parts kit including

    Furnace Photomultiplier tube Lamp
    Standard pilot burner
    Door seal
    Vent seals
    Vent actuator
    Exhaust fan
    Gas/air solenoid
    Fuse kit
    Computer system not included but can be supplied if requeste

     